产品介绍
"The sensors for free chlorine CCS14x is an amperometric, membrane-covered sensor for installation in the CCA250 assembly. Oxidising agents such as chlorine or anorganic chlorine compounds are used for disinfection of water. Their dosing must be carefully controlled to suit the application. Too low a concentration makes the degree of disinfection questionable. Too high a concentration can result in corrosion effects, impairment of taste or skin irritation.","Your benefits:"
特性
- Minimum flow rate for installation in the CCA250 flow assembly: 30 l/h
- Measurement almost independent of flow rate in the range above 30 l/h
- No zero point calibration necessary. This means complicated installation of an active carbon filter, as in open chlorine sensors, is not necessary.
- Measured values are not affected by conductivity fluctuation.
- The CCS140 sensor is ready for measurement after a polarisation time of approx. 30 ... 60 minutes. The CCS14x sensor requires 45 ... 90 min.
- Easy membrane replacement thanks to ready-made membrane head
- Recalibration intervals approx. 1 ... 4 months under constant operating conditions
- Back pressure up to 1 bar / 14.5 psi allowed at the outlet
CCS 140&141 CMOS图像传感器技术参数
点击查看 0 个替代产品
规格项 | 参数值 | 勾选搜索替代 |
特 性 | ||
测量参量 | Free Chlorine | |
响应时间 | 120 seconds | |
输出选项 | Analog Current | |
环 境 | ||
工作温度 | -4 to 140 F (-20 to 60 C) | |
更多技术信息 | ||
仪表类型 | Fixtured or Permanent | |
更多规格 | ||
高量程 | 20 mg/L | |
低量程 | 0.0100 mg/L | |
产品类别 | Water Quality Photometers and Colorimeters |
产品替代
找到 个替代产品
资料下载
CCS 140&141 CMOS图像传感器资源附件
文件名称 | 大小 | 操作 |
---|---|---|
TI00058CEN_1312 | 416.25 KB | 下载 |
CMOS图像传感器相关品牌
CMOS图像传感器同类产品