小程序
传感搜
传感圈

ValueTronics 309 毫欧表

分享
  • 后询价或查看价格

  • 309

  • 毫欧表

  • ValueTronics

  • 请咨询供应商

  • ValueTronics

联系销售
专家支持

特性

  • Accuracy: ± 1% of reading (when the test current is > 500μ / impedance in Ohms and > 30 mA)
  • Range: 0 to 100 Ohms
  • Measurement Technique: Successive approximation analog to digital conversion of sensed voltage and current
  • Measurement window: Either 2 or 3 full sine waves depending upon line frequency (e.g. 3 at 60 Hz, 2 at 50 Hz)
  • Sample rate: ~ 53 KHz, yielding ~ 2.6K voltage and 2.6K current samples per measurement window at 60 Hz, and ~ 2.1K voltage and 2.1K current samples/window at 50 Hz
  • Mathematical conversion to true rms voltage and current values, division independent of phase to yield impedance
  • Display derived from four layer stack averaging
  • Voltage is sensed at the probe heads by internal probe sense lines or, for Model 109 probes, by precision Kelvin sensing
  • Accuracy: ± 1% of reading ± 4 mA
  • Range: 0 to 144 Amps rms
  • Accuracy: ± 1% of reading ± 4 mA at transition points (every 2.5 Arms)
  • Range: 2.5 to 77.5 Amps rms (0 to 2.5 Arms implied) (Note 3, Graph 2)
  • Measurement Technique: Logical translation of digital current measurement
  • Accuracy: ± .1% (Digital Timer and Audio / Visual Alerts applicable to Model 309 or, if server driven, Model 308)
  • Range: 1 second to 59 minutes, 59 seconds (extended times possible under server software)
  • Internal Reference: 16.777 MHz (2e24) crystal controlled system clock
  • When enabled, sounds and flashes when any user specified test current or impedance limit is violated
  • Accuracy: Logically detected. The accuracy therefore exactly matches the accuracy of associated measurements
  • Continuous to test currents of 35 Amps
  • Derate linearly to 50% duty cycle at 60 Amps
  • Maximum Cycle Period: 35 to 60 Amps: 4 Minutes
  • Maximum allowed common mode voltage to ground: 50 Vrms, 71 Vpeak or dc
  • Maximum normal mode open circuit probe source voltage (at 132 Vrms line voltage): 8.6 Vrms, 12.1 Vpeak
  • 90 to 132 Volts rms
  • 48 to 66 Hertz sine wave
  • Maximum Power: 396 Watts (3.0 Arms at 132 Vrms)
  • Resistance: 100 milliohms ± 1%
  • Maximum continuous current: 17.5 Amps
  • There are no hardware calibration adjustments, and thus no hardware calibration is required or possible
  • Software calibration data is available, but it's normally unnecessary to modify this data
  • Complete the formal performance verification procedure annually or following any repair service
  • Designed and manufactured to meet or exceed UL 3111
  • Instruments manufactured before January 2008 were so listed by ETL
  • Operating Temperature Range: 5 to 40°C
  • Relative Humidity Range: 0 to 90% non-condensing
  • Height: 10.92 centimeters (4.30 inches)
  • Width: 21.15 centimeters (8.33 inches)
  • Depth: 26.0centimeters (10.24 inches)

"The 309 is a Milliohm Meter from Hypatia. When engineers need to collect information regarding electrical connections such as wires, conduits, or cables, they use a milliohm meter. This piece of electronic testing equipment measures the smallest resistance values. An engineer would use a milliohm meter instead of a traditional ohmmeter to document and log resistance amounts at the lower end of the scale (small values of ohms).","Additional Features:","Electrical","Impedance Measurement","Digital Current Measurement","Bar Graph Current Measurement","Digital Timer","Audio\/Visual Alerts","Operating Duty Cycle","Input Power Requirements","Check Resistor","Verification Schedule","Safety Certification","Environmental\\","Physical","Overall Dimensions","The Hypatia 309 is a High Current Sourcing Milliohm Meter."

产品替代

找到 个替代产品

资料下载

309 毫欧表资源附件
文件名称 大小 操作
Hypatia 308, 309 39.57 KB

电磁和电量传感器交流圈

约5圈友等你加入

    加载中···

    专家支持

    0 / 300

    选择图片

    确定提交

    提交成功

    系统自动将您的问题推送给专家解答

    查看专家解答
    请扫码进入传感圈小程序

    如何查看专家解答您的问题?

    1
    2
    3
    4